Test Methods: ASTM D445, D2532
Low Temperature Kinematic Viscosity Baths
• Conform to ASTM D445 and D2532
• Bath temperature stability and uniformity of ±0.05°C
• Mechanically refrigerated to -70°C
• Digital controller resolution 0.01°C
• Models with Working Temperature Range of +15ºC to –70ºC
Elevated Temperature Kinematic Viscosity Baths
Software Assistant for ASTM D445 Viscosity Testing

Lawler baths for low temperature kinematic viscosity conform to ASTM D445 and D2532 and related test method requirements for viscosity bath temperature stability and uniformity in the temperature range of from +15° to -75°C. A digital indicating PID temperature controller with 0.01°C resolution provides stability and uniformity of better than ±0.05°C as specified by ASTM D445 (when methyl alcohol or denatured ethyl alcohol are used as bath medium).

The bath jar is an evacuated, unsilvered Dewar flask. The jar is back-lit by a fluorescent light to facilitate reading of the meniscus in the viscometer tubes. A clear plastic viewing window protects the bath jar. Models have either 2 or 4 sample capacity.

Only ozone friendly, CFC and HCFC free refrigerant are used.

Model 77 is configured to accept 4 ea. 51 mm (2 inch) diameter holders. The 4 viscometer ports can be configured to any viscometer holder configuration, by special request. Model 77 temperature range is +15°C to -60°C.

Model 79 is identical to Model 77 but with the operating temperature range of +15° to -70°C.

Model 78 is identical to Model 77 but with the operating temperature range of +15° to -40°C.

Model 74 has a capacity for 2 viscometers with a temperature range of +15°C to -30°C.

Model 75 has a capacity fro 2 viscometers with a temperature range of +15°C to -40°C.

Model 77AV is identical to Model 77 but configured to accept either 1 or 2 Schott automated AVS or TC viscometer holders. It can also be configured to accept one or two automated viscometers by other manufacturers. Please specify the automated viscometer holder configuration and dimensions when ordering.

Model 79F is identical to Model 79 but with the additional configuration to accept 1 or 2 ASTM D2386 test glassware and stirrer (supplied separately) for performing freezing point of aviation fuels. Lowest operating temperature is -75°C.

Available Models
Model No.
Operating Temperature
Remarks
Cabinet Dimensions
Ship Weight lbs/cu.ft
74
+15  to  -30°C
2 Test Positions
18x22x34
(45x55x85)
250/15
75
+15° to -35°C
2 Test Positions
18x22x34
(45x55x85)
250/15
77
+15   to  -60°C
4 Tests Positions
22x22x26
(55x66x66)
300/17
77AV
+15   to  -60°C
2 Automated Viscometers
22x22x26
(55x66x66)
285/27
78
+15° to -40°C
4 Tests Positions
22x22x26
(55x66x66)
300/17
79
+15   to  -70°C
4 Tests Positions
22x22x26
(55x66x66)
300/17
79F
+15   to  -75°C
Viscosity by D445
and Freeze Point by D2386
22x22x26
(55x66x66)
300/17
KVS-1
Floor stand for above models

Specifications subject to change without notice.
Lawler Manufacturing Corporation © 2010