Model 321 series of oxidation stability baths conform to the requirements of EN-14122 test method, soon to be incorporated in ASTM D6751 specifications for B100 fatty acid methyl esters (FAME) for biodiesel blending.
Models 321 series employ multiple aluminum blocks configured to fit the specified test tube holding 3.0 gm of FAME (fatty acid methyl esters). When the sample reaches test temperature of 110°C filtered, dry air is sparged at 10 ± 1L/hr. to each test specimen.
The volatiles generated (organic acids) from this oxidation process are collected in a receiver of distilled or deionized water. The increase of acidity in the receiver is measured with time with an integral conductivity measuring device. The oxidation induction period is automatically calculated based on the break in the rate of increase (second derivative) of conductivity in the receiver.
Each position’s receiver has cap with an integral conductivity probe reporting data to the PC.
Model 317-8 is an economical liquid bath with a range of ambient to 150°C with 8 test positions.
|
321-2
|
2
|
Economical model, single block bath, supplied with operating software, glassware, connecting tubing, two conductivity probes, optional air filter/dryer and pump, PC not included.
|
|
321-4
|
4
|
Single block bath with 4 test positions, supplied with operating software, glassware, connecting tubing, 4 conductivity probes, optional air filter/dryer and pump, PC not included.
|
|
321-8
|
8
|
Fully equipped with air pump, filter, dryer, laptop PC, operating software, 2 independent aluminum heated blocks each with 4 test positions, 8 conductivity probes, all glassware and connecting tubing (for 8 positions).
|
|
321-16
|
16
|
Fully equipped with air pump, filter, dryer, laptop PC, operating software, 4 independent aluminum blocks each with 4 test positions, 16 conductivity probes, all glassware and connecting tubing (for 16 positions).
|
|
317-8
|
8
|
Economical liquid bath with flow meter, supplied with operating software, glassware, connecting tubing, conductivity probes, optional air filter/dryer and air pump. PC not included.
|
|
|
EN-10
|
Oxidation reaction vessel |
|
EN-11
|
Sealing cap for reaction vessel with air inlet tube, and outlet tubing
|
|
EN-12
|
Measuring vessel |
|
EN-13
|
Cap for measuring vessel with probe, inlet tube and vent |
|
EN-14
|
Certified temperature probe with digital display |
|
EN-15
|
Portable air pump with filter and molecular sieve dryer
|
|
EN-16
|
Molecular sieve dryer with filter assembly
|
|
EN-17
|
Silicone connecting hose, 50 ft.
|
|
EN-18
|
Pipette, 50 mL |
|
EN-19
|
Desiccator
|
|
EN-20
|
Drying oven to 150°C
|
|
EN-21
|
Optional float in tube flow meter with precision control valve |
|
EN-22
|
Pressurized house air setup for converting to instrument air |
|
|
|